Binder KBF PRO 720 Photostability Climat Chamber with ICH light modul
The Binder KBF PRO 720 standard climate chamber with extended temperature and humidity range is an ideal choice for photostability testing in accordance with the ICH Q1B guideline. The chamber can be equipped with a special ICH light module that provides standardized UV and visible light (VIS) exposure. The system is designed to enable fast and reliable completion of the three-day photostability test with minimal thermal impact. The KBF PRO 720 ensures a precise and stable environment, guaranteeing the reproducibility of test results in various industrial and research applications.
Package contents
Binder KBF PRO 720 Constant climate chamber with expanded temperature / humidity range
• Climate range: +10 °C to +50 °C and 10% to 80% RH
• Proven APT.line ™ preheating chamber technology
• Energy-efficient cooling with modern inverter cooling unit
• Cooling with climate-neutral R-290 refrigerant
• Door heating to prevent condensation
• Adjustable fan speed
• Humidity regulation with capacitive humidity sensor and responsive steam humidification
• Robust housing design with HIT sandwich technology
• Stainless-steel inner chamber
• Inner doors made of metal oxide-coated safety glass
• Two racks made of stainless steel with telescopic rails
• Large access port with silicone plug
• Robust twin casters, with brakes at the front
• Intuitive touchscreen controller with time-segment and real-time programming
• Ergonomically adjustable control terminal
• Signal triangle with light-up status display
• Internal data logger, measured values can be read out in open format via USB
• Class 3.3 independent temperature safety device (DIN 12880) with visual and acoustic alarms
• Unit self-test for comprehensive status analysis
• Data interfaces: Ethernet, USB
• BINDER Multi Management Software APT-COM™ basic edition
ICH light module without light dose control
The ICH light module is connected to the climate chamber via a data cable. This makes it possible to control the irradiation duration of the light cassettes individually with a time segment program via the program controller of the climate chamber.
Scope of delivery of the BINDER ICH light modules
• VIS light cassette with connection cable
• UVA light cassette with connection cable
• U-rails for mounting the light cassettes in the climate control cabinet
• Supply box with ballasts/power supply units
• Profile rails and bolts for attaching the supply box to the climate chamber
• Data cable
• Mains cable
• Operating instructions