The new Solid.Line KBF-S climate chamber is a custom-made solution for long-term testing and accelerated stability testing for pharmaceutical products in accordance with the ICH Q1A guideline. The corrosion-free and condensation-free interior also ensures exceptionally stable temperature and moisture levels, even when fully loaded.
Benefits
• Homogeneous climate conditions thanks to APT.line™ preheating chamber
• Standard-compliant testing according to ICH Q1A
• Precise temperature and humidity control
Important characteristics
• Temperature range: 0 °C to +70 °C
• Humidity range: 20% to 80% RH
• APT.line™ preheating chamber technology
• Humidity regulation with capacitive humidity sensor and steam humidification
• LCD to display temperature and humidity along with additional information and alarms
• Internal data logger, measured values can be read out in open format via USB
• Unit self-test for comprehensive status analysis
• Inner chamber made completely of stainless steel
• 2 stainless steel racks, incl. shelf supports
• Access port with silicone plug, 30 mm, left
• Robust twin casters with brakes, from 240 L
• Class 3.1 independent temperature safety device (DIN 12880) with visual and audible temperature alarm
• Computer interface: Ethernet
• Independent water supply via tank

The new Solid.Line KBF-S climate chamber is a custom-made solution for long-term testing and accelerated stability testing for pharmaceutical products in accordance with the ICH Q1A guideline. The corrosion-free and condensation-free interior also ensures exceptionally stable temperature and moisture levels, even when fully loaded.
Benefits
• Homogeneous climate conditions thanks to APT.line™ preheating chamber
• Standard-compliant testing according to ICH Q1A
• Precise temperature and humidity control
Important characteristics
• Temperature range: 0 °C to +70 °C
• Humidity range: 20% to 80% RH
• APT.line™ preheating chamber technology
• Humidity regulation with capacitive humidity sensor and steam humidification
• LCD to display temperature and humidity along with additional information and alarms
• Internal data logger, measured values can be read out in open format via USB
• Unit self-test for comprehensive status analysis
• Inner chamber made completely of stainless steel
• 2 stainless steel racks, incl. shelf supports
• Access port with silicone plug, 30 mm, left
• Robust twin casters with brakes, from 240 L
• Class 3.1 independent temperature safety device (DIN 12880) with visual and audible temperature alarm
• Computer interface: Ethernet
• Independent water supply via tank


Specifications for Binder  KBF-S 1020 UL standard climate chamber

Interior volume (liter) 1020
Internal dimension W x H x D (mm) 976x1250x836
Number of shelves (std./max.) 2/15
Housing dimensions W x H x D (mm) 1249x1925x1193
Rated voltage (V) 200…230