New Kühner LS-Z Kelvin+ Incubator shaker

%
Kühner Kelvin+ incubator chamber
Kuhner Kelvin+ can be placed on, or under, any table or lab bench. The incubator has a flat front and a space-saving door mechanism. The large window provides a good view of cultivation vessels. For static experiments, a fixed shelf may be placed above the shaking tray inside the chamber. The 40 mm feedthrough port provides enough space for tubing and cables needed for the cultivation process. An exterior side-mounted document holder adds convenience to the Kuhner Kelvin+.
The stainless-steel chamber of the Kuhner Kelvin+ is easily cleaned.
Technical data Kuhner Kelvin+
• Temperature distribution: +/- 0.5°C (at 37°C, 100 rpm)
• Principle of sensor: Pt-100
• Air circulation: 190 m3/h
• Illumination: LED
• Ambient temperature: 10 up to 35°C
• Operating menu in: DE, FR, EN, IT, ES
• Interface standard: CAN-Bus
• Tray, size: E (420 x 420mm)
• Control accuracy, absolute: <+/- 0.3°C
• Material: inner chamber stainless steel, outer chamber PUR (insulation), door polycarbonate
%
Kühner Kelvin+ shaking incubator
A compact sized shaker-incubator with an excellent temperature distribution
The Lab-Shaker LS-Z and the incubator Kuhner Kelvin+ combine approved Kuhner quality in a new modular system to provide maximum flexibility to our customers. The LS-Z, just like its predecessor LS-X, is a sturdy benchtop shaker designed for continuous and maintenance-free ope­ration. Our new Kuhner Kelvin+ is a compact sized incubator with an excellent temperature distribution.
Performance characteristics
• Compact footprint
• Hygienic design: easy cleaning
• Fast regeneration time due to stainless steel chamber
• Very good temperature distribution
• Large feedthrough port
• Kuhner TOM online measurement tool upgradable
Technical data - Lab-Shaker LS-Z
• Power consumption typical: approx. 25W
• Weight shaker: 46 kg
• Weight Power Cube: 6.5 kg
• Weight Controller: 0.6 kg
• Ambient temperature: -20 up to 80°C
• Ambient relative humitidy: up to 90 % (shaking unit)
• Operating menu in: DE, FR, EN, IT, ES
• Interface standard: CAN-Bus
• Tray, size
E (420 x 420mm)
EX (500 x 420mm)
F (800 x 420mm)
• Loading maximum: up to 25 kg
• Setting, digital: 1 rpm
• Control accuracy, absolute: +/- 0.5 rpm
• Timer: 1 s – 999 h
• Acceleration: controlled
• Active break: adjustable
• Stop on position: adjustable
• Shaking motion
orbital, Ø 12.5 mm* 20...500 rpm
orbital, Ø 25.0 mm* 20...400 rpm
orbital, Ø 50.0 mm* 20...300 rpm
• Material: Stainless steel cover
Options & Accessories
Interface, optional: USB, Ethernet, digital, analogue
IQ/OQ-Documentation: available
Dual table: available
Technical data Kuhner Kelvin+
• Temperature minimum: ambient + 3°C
• Temperature maximum: 60°C
• Temperature distribution: +/- 0.5°C (at 37°C, 100 rpm)
• Principle of sensor: Pt-100
• Power of heating: 400W
• Air circulation: 190 m3/h
• Power consumption typical: approx. 130W
• Gas volume: 130 L
• Weight shaker: 28 kg
• Illumination: LED
• Ambient temperature: 10 up to 35°C
• Operating menu in: DE, FR, EN, IT, ES
• Interface standard: CAN-Bus
• Tray, size: E (420 x 420mm)
• Setting, digital: 0.1°C
• Control accuracy, absolute: <+/- 0.3°C
• Timer: 1 s – 999 h (shaker only)
• Material: inner chamber stainless steel, outer chamber PUR (insulation), door polycarbonate
Options & Accessories
• Light shading plate
• Shelf
• Interface, optional
• Stacking adapter
• IQ/OQ-Documentation
• Dual table