Heidolph MR Hei-Connect magnetic stirrers with heating functions

NEW
Heidolph Hei-Connect magnetic stirrer with heating function with Pt 1000 temperature sensor
For comprehensive process documentation and reproducible results
• Precise setting options and monitoring of device parameters via the digital display
• With port for the optional temperature sensor Pt 1000 for precise values without temperature
• fluctuations
• All values can be easily read on the illuminated display even from a distance
• Illuminated buttons indicate activated functions
• With the control knob, the speed can be accurately set between 100 and 1,400 rpm with a speed
• accuracy of ±2 % and a maximum temperature of 300 °C
• To protect against overheating, an independent safety circuit switches the heating off if the set temperature is exceeded by 25 °C
• To protect the sample, the external sensor monitoring automatically switches off the heating
• if the sensor is not immersed in the medium
• The timer function allows you to define separate expiry times for the heating and rotation functions. At the end of the process an acoustic signal sounds
• Residual heat indicator effectively protects from burns
NEW
Heidolph Hei-Connect magnetic stirrer with heating function
For comprehensive process documentation and reproducible results
• Precise setting options and monitoring of device parameters via the digital display
• With port for the optional temperature sensor Pt 1000 for precise values without temperature
• fluctuations
• All values can be easily read on the illuminated display even from a distance
• Illuminated buttons indicate activated functions
• With the control knob, the speed can be accurately set between 100 and 1,400 rpm with a speed
• accuracy of ±2 % and a maximum temperature of 300 °C
• To protect against overheating, an independent safety circuit switches the heating off if the set temperature is exceeded by 25 °C
• To protect the sample, the external sensor monitoring automatically switches off the heating
• if the sensor is not immersed in the medium
• The timer function allows you to define separate expiry times for the heating and rotation functions. At the end of the process an acoustic signal sounds
• Residual heat indicator effectively protects from burns